Provides integrated measurement solutions including meters, algorithms, automation programs, toolboxes, and sample rods
|
Parameter Item
|
Specification
|
|
Operating Environment
|
Temperature: 1.4 K ~ 400 K
Pressure: 10⁵ ~ 10⁻⁵ Pa
Magnetic Field: 0 ~ 18 Tesla
|
|
Compatible Outer Diameter
|
dia-26 mm / dia-30 mm / dia-50 mm
|
|
Measurement Functions
|
AC magnetic susceptibility, AC susceptibility vs. temperature/frequency/magnetic field curve measurement, etc.
|
|
Applicable Sample Types
|
Bulk, thin film, powder samples of various geometries
|
|
Sample Limits
|
Mass: < 20 g
Radial dimension: < 3 mm
|
|
Measurement Range
|
Signal Range: 1 × 10⁻⁵ emu/Oe ~ 40 emu/Oe
|
|
Accuracy
|
Amplitude: ± 0.5%
Phase: ± 0.5 deg
|
|
Noise Floor
|
< 1 × 10⁻⁸ emu/Oe (1 s integration time)
|
|
Optional Compatible Accessories
|
Standard quartz sample rod, oxygen-free copper sample rod, electrically controlled magnetic sample rod, optically controlled magnetic sample rod, high-voltage VSM sample rod
|